Product Name: iCAM Alpha/Beta Air Monito

Product Description

CANBERRAÂ’s intelligent alpha/beta Continuous Air Monitor (iCAM) provides robust and reliable monitoring of airborne alpha and beta particulate activity in the workplace.
It acts as a simple alarming monitor for operators while measuring airborne activities in real time with the sophistication required to provide low false alarm rates and high protection levels. At the same time it provides automated facilities which assist supervisors to conduct detailed setup and operational overview.

The iCAM measures airborne alpha and beta particulate activity with radon/thoron alpha and beta background compensation. It also provides static or dynamic compensation of beta measurements for gamma background depending on the choice of detector type.

Air is drawn through the instrument by means of an external pump or vacuum system, and airborne particulate material is deposited on the removable card-mounted filter. The filter is monitored by a robust CANBERRA CAMS PIP (passivated ion-implanted planar silicon) radiation detector, which allows simultaneous measurement of both alpha and beta radioactivity in the material deposited on the filter. The air flow is measured directly and reported by the instrument. The design of the airflow system is optimized to ensure high air sampling efficiency and high transmission of particles to the filter. The iCAM has a high detection efficiency for both alpha and beta particles and provides a good sensitivity for low energy beta detection.

Measurement results are presented on the high-visibility graphical display on the instrumentÂ’s front panel. User setup and diagnostics are via the local keypad and the graphical display which provide a natural user interface. Results may also be reported via the RS-232 or RS-485 network connections to a local computer or plant surveillance system. Each instrument stores historical data, typically from the previous seven days, which may also be transmitted by network if required. The historical data includes not only activity readings, updated every five minutes in normal operation, but also alarms, filter changes and user interactions. Each event is logged with its date and time.

The iCAM Primary Alarm uses measurements taken over a user selectable interval of typically five minutes, and this data is updated as rolling average every 15 seconds. Alpha, beta and gamma (where applicable) measurement values are each tested against user-set alarm levels. There are three separate sets of three alarm levels, one set of three for each of the alpha, beta and gamma measurement variables. Additional defence against larger emissions of activity in a short time is provided by the Quick Alarm, for which decisions are made on the new data accumulated every 15 seconds.

Finally, iCAM also provides a Long Term Alarm (Chronic Activity Alarm). Here the lowest level alarm may be user-allocated to data accumulated over a long time interval, typically >30 minutes, so that the most sensitive measurements can be achieved. All the alarms are fully compensated for the backgrounds from gamma radiations and from radon/thoron daughter activities.

In addition, any alarms may be indicated locally by visual and audible signals, remotely by relay circuits, or centrally by connection to a suitable monitoring network. Alarm events are also archived.
Hardware

The iCAM instrument is packaged in a rugged steel enclosure which provides IP42 environmental protection and EMC screening. The front panel carries the local controls and indicators, including the high visibility graphical display, the local keypad and the various indicator lights. The RS-232 connector and the alarm sounder are positioned in the bottom third of the front panel. The air inlet and alarm beacons are mounted on top of the instrument.

When the power supply is switched on, the instrument will perform initialization tests and begin normal operation without any operator intervention. It will begin providing valid measurements within 15 seconds after starting normal operation. It will then update measurements at 15 second intervals, whatever the user-set counting interval.

Air is drawn into the instrument at a preset rate, within the range 15-60 L/min (0.5-2.1 ft3/min), typically about 37 L/min (1.3 ft3/min), and particulate material is deposited on a removable filter card. The instrument may be fitted with an adapter for the attachment of a 25 mm (1 in.) pipe. An airflow regulator valve is provided as an option, and the choice of air outlet position may be adopted from either of two connectors: one on top of and one on the bottom of the instrument. The airflow rate is measured directly on board and its value is used to calculate concentration values and integrated concentration values. Low and high flow alarm values can be set by the user. The total air flow since last reset is recorded and reported.
Detectors

A CANBERRA PIPS detector is used for detecting both alpha and beta activity. The single detector version provides static gamma compensation of beta measurements in low or steady gamma background conditions. Optionally, iCAM can come with a dual detector made from two similar detectors in a single housing, with the second detector located very close to the first. This second detector is not exposed to the airflow or filter and instead provides the data to make accurate dynamic compensation measurements for the effect of gamma backgrounds.

The measurement range of the instrument extends to over 500 kBq (13.5 µCi) of combined alpha and beta activity deposited on the filter. A background activity check measurement and display function is provided, which can be accessed via the front panel controls. This check may also be used to identify the presence of contamination within the monitor or to assess ambient gamma background values.

Features
Engineering: 
Removable, card-mounted filters
Automatic self-initialization and operation on power on
Electronic air flow rate measurement, with high and low flow alarms
Continuous total airflow recording and reporting
User-set alarms with local and remote reporting
Simple/automated checks for calibration, gain and spectrum shape
Software package for full iCAM configuration by local computer or via a network
Fault diagnostics show on the local display and across the network
RS-485 and RS-232 communication interfaces 
One-week data storage with detailed event log
Data is RS-232 and RS-485 transferable to Microsoft Excel
Thirty minute protection of full functionality by built-in battery backup
Five years protection of configuration memory
Measurement: 
Continuous monitoring of both alpha and beta activity in air 
Alpha MCA spectroscopy and separate beta discrimination 
Good low energy beta detection 
Auto-adaptive spectrometric compensation for radon/thoron 
Static gamma background compensation with single CAM PIPS detector
Fully compensated activity alarms 
Three activity levels for primary alarms 
Additional Quick and Long Alarm
Discharge and discharge rates for duct/stack monitoring
Options
Trolley mounted system 
Dual CAM PIPS detector version with dynamic gamma compensation 
Full-function area gamma alarm monitor capability using the optional G64 detector 
I/O expansion option provides RS-422 pulse train outputs (compatible with previous Harwell instruments), and also current loop input, optoelectronic outputs and other I/O 
Benefits
Engineering: 
A light and compact integrated unit, which is robust, simple to use and easy to maintain 
All the required facilities for alpha and beta monitoring, alarms, data logging, communications and setup 
Sensitive detection of alpha and beta particles plus intelligent adaptive algorithms, for fast, accurate alarms, resulting in high levels of protection and low false alarm rates 
Adaptive spectrometry algorithms and automated checks for high tolerance to variations in background conditions and extended time between filter changes 
Extra alarm facilities for improved management of safety

Display

A 240 x 128 pixel (100 x 80 mm) (4 x 3.15 in.) LED-backlit liquid crystal display panel presents operational information, and together with the panel controls, forms a natural interface for the user. It can show ongoing measurement data in various forms, including a live display of the accumulating alpha spectrum. The default is the measured activity levels in the pre-selected units. Alternatively it can be used, together with the panel controls, to show other information and to perform operational functions including calibration check, self tests, and viewing the archive.
Screens
The display shows the current measurement values in the display units selected via the /Display/Measurements soft key. 

After power-up, reboot or reset, the instrument will perform initialization tests and begin normal operation without operator intervention. With a filter card in place, the main screen is displayed, showing the activity readings and units (which may be selected from Bq, pCi, CPS, Bq/m3, pCi/L, Bq-hr/ m3, pCi-hr/m3, DAC, DAC-hr). When the instrument is configured as a duct monitor, the units may be in Bq or Bq/h. The units displayed are user configurable. All are measured. The rest of the display shows alarm/status messages, and time and date.

The lowest line of the main example display indicates the status of the four soft keys: the two righthand control keys provide service and maintenance functions and their use suspends normal monitoring. They are therefore accessed under password control. The two lefthand keys provide the operator with more information while normal operation continues uninterrupted.

Archive display may be selected by pressing the archive soft key. It allows the internal measurement and event archive to be inspected on the instrument display. The archive can hold enough entries of normal complexity, for seven daysÂ’ data. It includes activity measurements taken at (typically) five minute intervals, user actions such as reset, filter card release or calibration, and records all activity and fault alarms.

The table below shows the logical layout of the other screens. The check screens are protected by password and allow selected operators to conduct instrument checks, which can confirm satisfactory operation according to normal plant operating procedures.

The setup screens are password protected at a higher level. These allow supervisors or designated staff to configure the instrument parameters to match the facilityÂ’s operating procedures.
Alarms

iCAM provides visible indication for activity alarms by means of a flashing red beacon on the light stack and a two-tone audio alarm. An internal battery provides backup for the complete instrument operation, including alarm indication and reporting functions of the instrument for ~30 minutes after the loss of mains power. The light stack features a green light that indicates the measuring status of the instrument, i.e., it is in the “monitoring” condition, when the instrument is taking data. It provides similar information to the “monitoring” indicator on the front panel, but in a much more highly visible way. In particular, it flashes when the instrument enters the fault condition.

The local annunciation of activity alarms may be enabled or disabled (they are enabled at all levels by default). Irrespective of the annunciator assignment, the panel lights and message display screen also indicate alarm and fault conditions.

Relays 

The instrument has four relays, each providing two pairs of contacts, operated in a fail-safe mode (i.e., relays are normally energized and contacts normally closed). The relays are allocated by the user and the alarm conditions which actuate the alarm relays are user-configurable. Normally one relay is assigned to fault reporting. The alarm relays also indicate an alarm condition in event of loss of power to the instrument.

Alarm conditions may be latching (default) or non-latching. Latched alarms persist until cleared by user intervention; non-latching alarms automatically reset if the alarm condition no longer exists. Indication of an alarm condition ceases on operation of the reset control. If the alarm condition persists, indication of the alarm will recur as soon as it is recognized.
Activity/Dose rate Alarms

Three alarm levels may be defined for each measurement type; i.e., three alpha activity alarm levels, three beta activity alarm levels, and three levels for gamma dose rate, if appropriate.
When an activity alarm occurs: 
The cause of alarm (alpha, beta, gamma) and the alarm level appear on the display.
Alarm relays will open or close as programmed for the specific alarm condition. 
A status report may be obtained via the network port which identifies the alarm condition and cause. 
The alarm indicator is illuminated and an audio alert will sound. 
An alarm entry is added to the archive.
Alpha/Beta Activity Alarms
The iCAM provides three types of measurement alarm facility for alpha and beta activity. These can be selected to run at the same time. 

Primary Alarm:
This is the alarm used to routinely define the required level of working protection in a facility with optimum time response. It uses the primary measurement values, which are computed on data accumulated over a userselected averaging interval, typically five minutes, with the data updated every 15 seconds. Alpha and beta measurement values are each tested against userset alarm levels (separate sets of three alarm levels for alpha and beta).

Quick Alarm:
To provide rapid warning of emissions of activity in a short time, iCAM also provides a Quick Alarm, made on the data accumulated in the last 15 seconds. Separate single alarms (in cps only) are set at higher levels than the primary alarm levels for both the alpha and beta counting rates.

Long Term Alarm:
When time response is less of a priority, the Long Term Alarm allows the most sensitive and most accurate monitoring, recording workplace activities at the lowest levels.

This alarm uses measurements accumulated over a considerably longer averaging time, (e.g., 30–60 minutes), than the other alarms, but is still updated and tested every 15 seconds. The lowest level of the three primary alarm levels is dedicated to the Long Term Alarm test.
External Dose rate Alarms 

The external dose rate measured by the optional G64 gamma monitor detector is computed and tested against the three user-set gamma alarm levels at intervals of one second.

Self Check and Equipment Failure Alarm

The instrument will recognize and alarm on any of the following fault (equipment failure) conditions, each of which can be individually enabled or disabled: 
Low air flow or high air flow (user settable limits).
Instrument configuration (clock, parameter store, archive).
Low differential pressure across the filter. 
Detection fault – low count rate (alpha/beta, internal gamma, external gamma detector).
AC power failure.
Battery and power supply failure.
When an equipment failure alarm occurs: 
The green beacon on the light stack flashes and the audio alarm sounds a continuous tone. 
If configured, the equipment failure relay de-energizes to send a signal to external equipment.
For some faults: 
The equipment failure indicator is illuminated.
The cause of failure appears on the integral display.
A status report may be requested via the serial communications port, which identifies the equipment failure alarm condition and cause. 
An equipment failure alarm entry is added to the archive.
Data Archive 

The iCAM stores historical data at user-definable intervals. Measurement data, records of the filter changes, alarms and user actions are recorded. Each entry in the archive is marked with the date and time when it was created. There is adequate storage to retain a weekÂ’s data (taken on the standard five minute primary alarm interval). Battery protection allows the internal archive data to be retained for five years after loss of mains power.
Air Flow Rate Monitoring 

The iCAM electronically measures the air flow rate through the instrument, and the measured value is used in activity calculations. The air flow rate will normally be within the working range of 15-60 L/min (0.5-2.1 ft3/min), and is usually adjusted to be about 37 L/min (1.3 ft3/min) using the (optional) air flow regulator valve. Via the input/output expansion board option, iCAM can provide a 4-20 mA input channel for connection of an external duct flow measurement system. The iCAM can use this signal to calculate the activity discharged and the discharge rate of activity from the duct. 
Calibration 

iCAM provides facilities to assist and partially automate calibration and the routine checks required by plant procedures to confirm proper operation of such instruments. These facilities are accessed under password control via the check button on the main screen display.
Source Calibration and Testing 

Activity measurement calibration checks are performed using test sources in place of the filter card. The sources are mounted on stainless steel carriers of similar dimensions to the usual filter cards, allowing easy insertion. The recommended calibration test sources are manufactured by AEA Technology QSA, with the activity homogeneously incorporated in a thin anodized layer on an aluminium disc. Self-absorption is small and the sources are classified as thin sources according to ISO 2919.

Company Details

For over 35 years, the name Canberra has been known for world class nuclear measurement equipment and services. Now... meet the new Canberra.

... more

More Products of this Company: ACHNP97 & ACHNA98 Industrial Charge Amplifiers for Helium-3 Detectors, ADM606M Multi-purpose Radiation Monitor, AFT Research Amplifier - Model 2025, Alpha Analyst Integrated Alpha Spectrometer, Alpha Beta Environmental Continuous Air Monitor (ECAM), Alpha Sentry CAM, Alpha Spectrometers and Accessories, AMP/TSCA - Model 2015A, Arrangement of 2 or more Germanium Detectors, BARA 31 Gamma Alarm Beacon, Bin/Power Supply - Model 2000/2100, CAM110FF Series Continuous Air Monitors, Charge Amplifiers for Bonner Multisphere Systems, Charge Amplifiers for Neutron Scattering, Digital Signal Processor - Model 2060, Eclipse - Control and Analysis Software for S5XLB and LB5500 Systems, Edge/Crossover Timing Single Channel Analyzer - Model 2037A, Front Opening Split-Top Shield - Model 767, G64 Gamma Monitor, Gamma Analyst Integrated Gamma Spectromete, Ge Clover Detectors, Ge Encapsulated Detectors, Ge Segmented Coaxial Detectors, Geiger Mueller Detectors - GP Series, Geiger-Mueller (GM) Tubes, ICB Programmable Digital Signal Processor - Model 9660, iMatic Gas-less Automatic Alpha/Beta Counting System, iSolo Gas-less Manual Alpha / Beta Counting System, LB4100 Multi-Detector Low Background Alpha/Beta Counting System, LB5500 Large Area Automatic Low Background Alpha/Beta Counting System, Linear Ratemeter - Model 2081, Model 512 - Dual Counter/Timer, Model 8224 Analog Multiplexer, Model 8715 - Analog-to-Digital Converte, Model 9635 - ICB Programmable ADC, Model Calibrator/M Semi-Automatic Calibration and Check Bench, Networked Continuous Air Monitor (NetCam), Neutron Detectors - NP Series, PGM102 Portable Gas Monitor, Photomultiplier Tube Base/Preamplifier - Model 2007/2007P, Scintillation Detectors - Model 802, Series 5 APC Console Based Automatic Alpha/Beta/Gamma Counting Systems, Series 5E Console-Based Automatic Low Background Alpha/Beta Counting System, Series 5XLB Automatic Low Background Alpha/Beta Counting System, Si(Li) Charged Particle Detectors with Transmission Mounts, Si(Li) Segmented X-ray Detectors, Silicon (Li) Detector Systems, Spectroscopy Amplifier - Model 2026, Standard PIPS Detectors, TAM Series Tritium Air Monitor, X-PIPS Detector
Related Products: 265A Photomultiplier Base, 414A Fast Coincidence, 427A Delay Amplifier, 439 Digital Current Integrator, 460 Delay Line Amplifier, 474 Timing Amplifier, 533 DUAL SUM AND INVERT AMPLIFIER, 542 Linear Gate and Stretcher, 570 AMPLIFIER, 572A AMPLIFIER, 590A AMPLIFIER AND TIMING SINGLE CHANNEL ANALYZER, 671 SPECTROSCOPY AMPLIFIER, 673 Spectroscopy Amplifier and Gated Integrato, 863 QUAD TIMING FILTER AMPLIFIER, 9302 AMPLIFIER DISCRIMINATOR, 9327 1-GHz Amplifier and Timing Discriminator, ACHNP97 & ACHNA98 Industrial Charge Amplifiers for Helium-3 Detectors, ADM606M Multi-purpose Radiation Monitor, AFT Research Amplifier - Model 2025, ALPHA (NORM) CONTAMINATION IN PIPES, TOOLS AND OBJECTS, Alpha Analyst Integrated Alpha Spectrometer, Alpha Beta Environmental Continuous Air Monitor (ECAM), Alpha Sentry CAM, Alpha Spectrometers and Accessories, ALPHA SURFACE CONTAMINATION