Product Name: QE65000 Scientific-grade Spectrometer
Product Description
The QE65000 Scientific-grade Spectrometer is a novel combination of detector, optical bench and electronics technologies that provides users with a remarkably sensitive system for demanding low-light level applications such as fluorescence, DNA sequencing, astronomy and Raman spectroscopy. The QE65000 can achieve up to 90% quantum efficiency with high signal-to-noise and rapid signal processing speed. The QE65000 Scientific-grade Spectrometer is available for $10499.
Quantum Efficiency to 90%
The QE65000 is the most sensitive spectrometer we have ever developed. The Hamamatsu FFT-CCD detector used in the QE65000 provides 90% quantum efficiency (defined as how efficiently a photon is converted to a photo-electron). Most of our other detectors are linear CCDs but with this “2D” area detector, we can bin (or sum) a vertical row of pixels, which offers significant improvement in the signal-to-noise ratio (>1000:1) and signal processing speed of the detector compared with a linear CCD, where signals are digitally added by an external circuit.
Increased System Sensitivity
In our spectrometers with linear CCDs, the slit’s width, not its height, regulates the amount of light entering the bench because linear CCDs cannot efficiently collect the light from the entire height of the slit. But in the QE65000, the 2D area detector can better take advantage of the height of the slit and the additional light, greatly improving system sensitivity.
Excellent UV Response
Because the detector in the QE65000 is back-thinned (or back-illuminated), it has great native response in the UV and does not require the additional coatings that we typically apply to other detectors for UV applications.
Demanding Low Light-level Applications
The QE65000 Spectrometer is a great option for low-light level applications including fluorescence, Raman spectroscopy, DNA sequencing, astronomy and thin-film reflectivity. The TE-cooled (down to -15 °C) detector features low noise and low dark signal, which enables low-light-level detection and long integration times (analogous to the shutter speed of a camera) from 8 milliseconds to 15 minutes, with virtually no spectral distortion.
An example of a fluorescence setup with a QE65000 uses the PX-2 Pulsed Xenon Lamp as an excitation source and the CUV-ALL Cuvette Holder for samples. An optical fiber delivers excitation light to the sample holder and read light to the spectrometer. A filter, such as one of our Linear Variable Filters would block excitation light from entering the spectrometer
Onboard Programming
The QE65000 also has an onboard programmable micro-controller for controlling the spectrometer and accessories. You have access to 10 user-programmable digital inputs/outputs and a pulse generator for triggering other devices. You can use the I/Os to flash a lamp, stop/start a process, and send a message/alarm during the spectrometer’s integration period. The spectrometer’s operating parameters can be controlled through software. In fact, wavelength calibration coefficients unique to each spectrometer are programmed into a memory chip right on the spectrometer.
Plug-and-Play USB Operation
The QE65000 interfaces to a computer via USB 2.0 or RS-232 serial port. Data unique to each spectrometer are programmed into a memory chip on the QE65000; our SpectraSuite Spectroscopy Operating Software reads these values for easy setup and hot swapping among computers, whether they run on Linux, Mac or Windows operating systems.
Optical Bench Options
At Ocean Optics, we understand that your work is unique and that you want a spectrometer to fit your specific needs. With the QE65000 Spectrometer, you can configure the bench for your application. Choose QE65000 Optical Bench Options such as entrance aperture size, detector accessories, filters a grating and more. Our application scientists will help you determine which configuration will best work for you.
Installation and Operation Manual
Our QE65000 Manual provides you with instructions for setting up, calibrating and performing experiments with your QE65000 spectrometer.
Pricing
Specifications
|
Physical
|
| Dimensions: |
182 mm x 110 mm x 47 mm |
| Weight: |
1050 grams |
| Detector |
|
Detector:
|
Hamamatsu S7031-1006
|
| Detector range: |
200-1100 nm |
|
Pixels:
|
1024 x 58 (1044 x 64 total pixels); 24.6 µm square size
|
|
Pixel size:
|
24.576 μm square size
|
|
Pixel well depth:
|
300,000 electrons/well; ~ 1.5 million electrons/column
|
| Sensitivity: |
22 electrons/count for all
wavelengths; 250 nm: 26 photons/count |
| Quantum
efficiency: |
90% peak |
| Optical
Bench |
| Design: |
f/4, Symmetrical crossed
Czerny-Turner |
| Focal length: |
101.6 mm input and output |
| Entrance
aperture: |
5, 10, 25, 50, 100 or 200 µm
wide slits or fiber (no slit) |
| Grating options: |
14 different grating
options, UV through Shortwave NIR |
| HC-1 grating
option: |
provides 200-950 nm range |
| Detector
collection lens option: |
None |
| OFLV filter
options: |
OFLV-QE |
| Other bench
filter options: |
Longpass OF-1 filters |
| Collimating and
focusing mirrors: |
Standard only |
| UV enhanced
window: |
No |
| Fiber optic
connector: |
SMA 905 to 0.22 numerical
aperture single-strand optical fiber |
| Spectroscopic |
| Wavelength
range: |
Grating dependent |
| Optical
resolution: |
~0.14-7.7 nm FWHM |
| Signal-to-noise
ratio: |
1000:1 (at full signal) |
| A/D resolution: |
16 bit |
| Dark noise: |
3 RMS counts |
| Dynamic range: |
7.5 x 109
(system), 25000:1 for a single acquisition |
| Integration
time: |
8 ms to 15 minutes |
| Stray light: |
<0.08% at 600 nm; 0.4% at
435 nm |
| Corrected
linearity: |
>99.8% |
| Electronics |
| Power
consumption: |
500 mA @ 5 VDC (no TE
cooling); 3.5 A @ 5 VDC (with TE cooling) |
| Data transfer
speed: |
Full scans to memory every 7
ms with USB 2.0 port, 18 ms with USB1.1 port, 300 ms with serial port |
| Inputs/Outputs: |
10 onboard digital
user-programmable GPIOs (general purpose inputs/outputs) |
| Analog channels: |
No |
| Auto nulling: |
Yes |
| Breakout box
compatibility: |
Yes, HR4-BREAKOUT |
| Trigger modes: |
4 modes |
| Strobe
functions: |
No |
| Gated delay
feature: |
Yes |
| Connector: |
30-pin connector |
| Power-up time: |
<5 seconds |
|
Dark current:
|
4000 e-/pixel/sec @ 25 ºC; 200 e-/pixel/sec @ 0 ºC
|
| Computer |
| Operating
systems: |
Windows 98/Me/2000/XP, Mac
OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
| Computer
interfaces: |
USB 2.0 @ 480 Mbps; RS-232
(2-wire) @ 115.2 K baud |
| Peripheral
interfaces: |
SPI (3-wire); I2C
inter-integrated circuit |
| Temperature
and Thermoelectric (TE) Cooling |
| Temperature
limits: |
-15 ºC to 50.0 ºC; no
condensation |
| Set point: |
Software controlled; lowest
set point is 40 ºC below ambient, to -15 ºC |
| Stability |
+/-0.1 ºC of set
temperature in <2 minutes |
Company Details
Ocean Optics is a leading supplier of solutions for optical sensing – fundamental methods of measuring and interpreting the interaction of light with matter. We enable diverse applications in medical and biological research, environmental... more
| More Products of this Company: |
CHEM4-UV-FIBER Spectrophotometer, CHEM4-UV-VIS Spectrophotometer, CHEM4-VIS-FIBER Spectrophotometer, CHEM4-VIS-NIR Spectrophotometer, DTS-series MEMS-based NIR Spectrometers, Fiber Optic pH Sensors, HR2000+ High-resolution Spectrometer, HR2000+CG Spectrometer, HR2000CG-UV-NIR Spectrometer, HR4000 High-resolution Spectrometer, HR4000CG-UV-NIR Broadband Spectrometer, Hydrorad Environmental Radiometer, LIBS Broadband Spectrometer System, LIBS-ELITE Spectrometer System, NanoCalc Thin Film Reflectometry System, NIR-512 Near-infrared Spectrometer, NIR256-2.1 NIR Spectrometer and NIR256-2.5 Extended-range NIR Spectrometer, OOISensors Software for Oxygen Measurements, Oxygen Sensor, Phazir Handheld NIR Material Analyzer, Phenol Red pH Test Kit, PlasCalc Plasma Monitoring System, QE65000 Configured for Fluorescence, QE65000 Scientific-grade Spectrometer, Raman Systems R-3000 Spectrometers, RSL-1 Handheld Raman System, S1024DW Deep Well Spectrometer, SpecEl Ellipsometer System, USB-650 Red Tide Spectrometer, USB2000+ Miniature Fiber Optic Spectrometer, USB2000-FLG Gated Fluorescence Spectrometer, USB4000 Miniature Fiber Optic Spectrometer, USB4000-FL Fluorescence Spectrometer, USB4000-FL-450 Spectrofluorometer, USB4000-UV-VIS Spectrometer, USB4000-VIS-NIR Spectrometer |
| Related Products: |
74-series Lens Fixtures, 84-series Lens Fixture, 84-series Lens Fixture, Absorbing Glass Optical Filters, ACH-CUV-VAR Adjustable Collimating Lens and Cuvette Holder, ACH-CUV-VAR Adjustable Collimating Lens and Cuvette Holder, Adjustable Collimating Lens Holder, Broadband 200-1100 nm, CG-series Spectrometers, Cell for Capillary Electrophoresis, CFV-series FluoroVettes, CHEM4-UV-FIBER Spectrophotometer, CHEM4-UV-FIBER Spectrophotometer, CHEM4-UV-VIS Spectrophotometer, CHEM4-VIS-FIBER Spectrophotometer, CHEM4-VIS-FIBER Spectrophotometer, CHEM4-VIS-NIR Spectrophotometer, CHEMUSB4-UV-VIS Spectrophotometer, CHEMUSB4-VIS-NIR Spectrophotometer, Cosine Correctors for Emission Collection, CUV-ALL-UV 4-way Cuvette Holder for 1-cm Cuvettes, CUV-UV Holder for 1-cm Cuvettes, CUV-VAR Variable Pathlength Cuvette Holder, Cuvette Holder for 10-cm Cuvettes, CVD-UV and CVD-VIS Disposable Cuvettes, Diffuse Reflectance Standards |
|