Product Name: CHEM4-UV-FIBER Spectrophotometer
Product Description
The CHEM4-UV-FIBER covers the 200-885 nm range, has optical resolution of 1.0
nm (FWHM) and captures 3648 wavelengths in <1 second. The system is ideal for
relative irradiance and emission measurements.
The CHEM4-UV-FIBER combines a USB4000 Spectrometer with modular accessories
including an ISS-UV-VIS
Integrated Sampling System, a 300
µm solarization-resistant optical fiber and a 1-cm cuvette holder for
absorbance measurements.
The advantage of an optical fiber based system is flexibility. Add
accessories including reflectance
probes and dip probes for even more measurement options.
The CHEM4-UV-FIBER is available for $4199.
Features
Fully integrated, preconfigured system covers the 200-885 nm range at 1.0 nm (FWHM)
See changes as small at 0.1 Absorbance Units
Small-footprint system ideal for the laboratory and the classroom
Flexible optical fiber-based system; easily add accessories
Integrated deuterium tungsten halogen light source and cuvette holder
USB 2.0 interface for fast, hassle-free spectrometer-to-PC connection
Light Source and Sample Holder
The CHEM4-UV-FIBER setup uses the ISS-UV-VIS
Integrated Sampling System, a combination RF deuterium source with a
tungsten bulb in a housing connected to a holder for 1-cm cuvettes. The
ISS-UV-VIS attaches to the spectrometer via a 300 µm solarization-resistant
optical fiber. The advantage of this design is that the light source can be
easily coupled to other fiber optic accessories such as reflection probes or
transmission dip probes.
Programmable Microcontroller
The CHEM4-UV-FIBER uses our USB4000
Spectrometer. The USB4000 features an onboard programmable microcontroller
that provides you with considerable flexibility in controlling the spectrometer
and accessories. Through a 22-pin connector, you can implement all operating
parameters in the software: control light sources, create processes, and
retrieve information on external objects. You have access to 8 user-programmable
digital Inputs/Outputs for interfacing to other equipment. (Contact
an Applications Scientist regarding what you'll need to program the GPIOs.)
Modular Design
Ocean Optics offers complete line of fiber
optic sampling accessories and teaching
tools to further enhance your CHEM4-UV-FIBER system. Most accessories have
SMA 905 connectors for application flexibility; changing the sampling system is
as easy as unscrewing a connector and adding new components or accessories.
Plug-and-Play Operation
The CHEM4-UV-FIBER interfaces to a PC, PLC or other embedded controllers via USB 2.0 or RS-232 serial port. When using the serial port, the system requires a single 5-volt power supply (not included). Data unique to each spectrometer is programmed into a memory chip on the USB4000; our spectrometer operating software reads these values for easy setup and hot swapping among PCs.
Serial port operation tips: When using the serial port, the CHEM4-UV-FIBER requires a single 5-volt power supply (not included). Also, SpectraSuite Spectroscopy Software does not support spectrometer operation via the serial port. However, using the included Serial Port Command Set, you can write your own software that makes serial port operation possible.
Specifications
|
Physical
|
| Spectrometer
dimensions: |
89.1
mm x 63.3 mm x 34.4 mm |
| Spectrometer
weight: |
190
grams |
| Light
source dimensions: |
198
mm x 104.9 mm x 40.9 mm |
| Light
source weight: |
400
g |
| Detector
Specifications |
|
Detector:
|
Toshiba TCD1304AP Linear CCD
array
|
|
Pixels:
|
3648 pixels
|
|
Pixel size:
|
8 μm x 200 μm
|
|
Pixel well depth:
|
100,000 electrons
|
| Signal-to-noise
ratio: |
300:1
(at full signal) |
| A/D
resolution: |
16
bit |
| Corrected
linearity: |
>99.8% |
| Sensitivity: |
130
photons/count at 400 nm; 60 photons/count at 600 nm |
| Optical
Bench |
| Design: |
f/4,
Asymmetrical crossed Czerny-Turner |
| Focal
length: |
42
mm input; 68 mm output |
| Entrance
aperture: |
25
µm wide slit |
| Grating: |
Grating
#1 |
| OFLV
filter: |
OFLV-200-850 |
| UV
enhanced window: |
Yes,
UV4 |
| Fiber
optic connector: |
SMA
905 to 0.22 numerical aperture single-strand optical fiber |
| Spectroscopic |
| Wavelength
range: |
200-885
nm
(The actual range
typically exceeds what's listed here. The range listed is where you will
see changes as small as 0.01 absorbance units.) |
| Optical
resolution: |
1.0
nm FWHM |
| Signal-to-noise
ratio: |
300:1
(at full signal) |
| A/D
resolution: |
16
bit |
| Dark
noise: |
50
RMS counts |
| Integration
time: |
3.8
ms to 10 seconds |
| Dynamic
range: |
2
x 108 (system), 1300:1 for a single acquisition |
| Stray
light: |
<0.05%
at 600 nm; 0.10% at 435 nm |
| Light
Source/Sample Holder |
| Power
consumption: |
420
mA @ 12 VDC; external power supply (included) |
| Pathlength: |
1
cm |
| Cuvette
shape: |
Square |
| Light
source: |
Deuterium
tungsten |
| Z
dimension: |
15
mm |
| Bulb
life (hours): |
800
deuterium; 2,000 tungsten |
| Time
to stabilized output: |
~30
minutes |
| Optical
Fiber |
| Optical
fiber: |
P300-1-SR;
300 μm diameter, solarization resistant |
| Electronics |
| Power
consumption: |
250
mA @ 5 VDC |
| Data
transfer speed: |
Full
scans to memory every 5 ms with USB 2.0 port, 18 ms with USB 1.1 port |
| Inputs/Outputs: |
Yes,
8 onboard digital user-programmable GPIOs |
| Analog
channels: |
No |
| Auto
nulling: |
Yes |
| Breakout
box compatible: |
Yes,
HR4-BREAKOUT |
| Trigger
modes: |
4
modes |
| Strobe
functions: |
Yes |
| Connector: |
22-pin
connector |
| Computer |
| Operating
systems: |
Windows
98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS
with serial port |
| Computer
interfaces: |
USB
2.0 @ 480 Mbps (USB1.1 compatible); RS-232 (2-wire) @ 115.2 K baud |
| Peripheral
interfaces: |
SPI
(3-wire); I2C inter-integrated circuit |
| Pricing |
| Retail: |
$4199 |
Company Details
Ocean Optics is a leading supplier of solutions for optical sensing – fundamental methods of measuring and interpreting the interaction of light with matter. We enable diverse applications in medical and biological research, environmental... more
| More Products of this Company: |
CHEM4-UV-VIS Spectrophotometer, CHEM4-VIS-FIBER Spectrophotometer, CHEM4-VIS-NIR Spectrophotometer, DTS-series MEMS-based NIR Spectrometers, Fiber Optic pH Sensors, HR2000+ High-resolution Spectrometer, HR2000+CG Spectrometer, HR2000CG-UV-NIR Spectrometer, HR4000 High-resolution Spectrometer, HR4000CG-UV-NIR Broadband Spectrometer, Hydrorad Environmental Radiometer, LIBS Broadband Spectrometer System, LIBS-ELITE Spectrometer System, NanoCalc Thin Film Reflectometry System, NIR-512 Near-infrared Spectrometer, NIR256-2.1 NIR Spectrometer and NIR256-2.5 Extended-range NIR Spectrometer, OOISensors Software for Oxygen Measurements, Oxygen Sensor, Phazir Handheld NIR Material Analyzer, Phenol Red pH Test Kit, PlasCalc Plasma Monitoring System, QE65000 Configured for Fluorescence, QE65000 Scientific-grade Spectrometer, QE65000 Scientific-grade Spectrometer, Raman Systems R-3000 Spectrometers, RSL-1 Handheld Raman System, S1024DW Deep Well Spectrometer, SpecEl Ellipsometer System, USB-650 Red Tide Spectrometer, USB2000+ Miniature Fiber Optic Spectrometer, USB2000-FLG Gated Fluorescence Spectrometer, USB4000 Miniature Fiber Optic Spectrometer, USB4000-FL Fluorescence Spectrometer, USB4000-FL-450 Spectrofluorometer, USB4000-UV-VIS Spectrometer, USB4000-VIS-NIR Spectrometer |
| Related Products: |
74-series Lens Fixtures, 84-series Lens Fixture, 84-series Lens Fixture, Broadband 200-1100 nm, CG-series Spectrometers, CHEM4-UV-FIBER Spectrophotometer, CHEM4-UV-VIS Spectrophotometer, CHEM4-VIS-FIBER Spectrophotometer, CHEM4-VIS-FIBER Spectrophotometer, CHEM4-VIS-NIR Spectrophotometer, CHEMUSB4-UV-VIS Spectrophotometer, CHEMUSB4-VIS-NIR Spectrophotometer, Cosine Correctors for Emission Collection, CUV-ALL-UV 4-way Cuvette Holder for 1-cm Cuvettes, CUV-UV Holder for 1-cm Cuvettes, CUV-VAR Variable Pathlength Cuvette Holder, Cuvette Holder for 10-cm Cuvettes, DTS-series MEMS-based NIR Spectrometers, FHS-UV In-Line Filter Holder, FIA-SIA-LOV Sequential Injection Analysis System, Fiber Optic pH Sensors, Fiber Optic Scanning Monochromator, Fiber Optic Variable Attenuator, Flow Cell Kit for Flow Injection Analysis, Flow Cell Kit for Flow Injection Analysis, Fluorescence Flow Cell |
|